21 CFR Part 11 explained: electronic records and signatures

21 CFR Part 11 is the FDA rule that decides when an electronic record or electronic signature is as trustworthy as its paper equivalent. If you keep quality or clinical records electronically — and almost every device company now does — Part 11 sets the conditions those records must meet.

The core requirements

What the rule actually asks for

Trustworthy, reliable records

Records must be accurate and retrievable throughout the retention period. That means controlled storage, protection from unauthorised change, and the ability to produce readable copies for FDA inspection.

Audit trails

Computer-generated, time-stamped audit trails must record creation, modification and deletion of records — who changed what and when — without obscuring earlier values.

Electronic signatures

Signatures must uniquely identify the signer, be applied only by the signer, and include the printed name, date and time and the meaning of the signature (reviewed, approved, and so on).

System controls

Access control, authority checks and device limits keep the right people in the right roles. Systems must also be validated for their intended use — the FDA's 2003 guidance scopes validation to the risk the records carry.

Making it real

A practical path to Part 11 compliance

  1. 1Inventory the records you keep electronically and decide which are Part 11 records.
  2. 2Prefer vendors whose systems already provide audit trails, unique accounts and signature manifestation — then validate your configuration.
  3. 3Write the procedures Part 11 expects: account management, signature use, record retention and copies for inspection.
  4. 4Train users so signatures stay personal, and review audit trails periodically rather than only when something goes wrong.
  5. 5Reassess when systems change — upgrades, migrations and new integrations all need a validation look.

Where Qevatrix fits

Part 11 controls built into the quality system

QualityOS applies Part 11-style controls across the quality record: unique named accounts, audit trails on every record, and electronic signatures that carry the signer's name, time and signature meaning — with nobody able to approve their own work.

Does buying software establish Part 11 compliance?

No. The organization must assess applicable electronic records, intended use, procedures and configured controls. Technical features such as audit trails and signatures support that work but do not replace it. Document the assessment and validation evidence, including how records can be retrieved and inspected over their retention period.

What does a practical example look like?

Illustrative example: a team moves an approval process from paper to an electronic system. It identifies the applicable record requirements, tests configured access and signature behavior, and confirms that a readable copy includes the signature meaning and history. A vendor feature list alone does not demonstrate the configured process is fit for use.

What should the reviewer check before proceeding?

  • Determine scope using applicable predicate rules and FDA guidance.
  • Test identity, authority, signatures and readable record copies.
  • Keep validation, operating procedures and training evidence under control.

Where can you check the original requirements?

Read the primary sources and confirm their applicability to your device and market. This example is illustrative, not a classification decision, clinical conclusion or approval.

How do you define the electronic-record boundary?

Start with a record inventory rather than a list of software features. For each record type, identify the underlying requirement to create or retain it, the authoritative copy, the people who can change it and the expected retention period. Include attachments, approval history and exported copies. A spreadsheet used only for temporary calculations may have a different role from the approved result retained as evidence; document that distinction instead of declaring all spreadsheets outside scope.

Follow one record through creation, review, signature, correction and retrieval. This exposes boundaries that a screen demonstration can miss, such as an attachment stored elsewhere or a migration that carries the final value but loses earlier changes. Evaluate closed and open system considerations against the actual access arrangements. An account password is one control, not a complete assessment of the record lifecycle.

What should an intended-use test demonstrate?

Illustrative test design: use three fictional roles—author, reviewer and unauthorized viewer—to exercise a controlled approval. Define the expected result before testing. Check an allowed action and a denied action, signature identity and meaning, and the readable record copy. The number of roles is an example, not a required validation sample size. Record configuration, expected outcomes, actual outcomes and deviations so another reviewer can understand what was tested.

Also test correction and retrieval after approval. Confirm that earlier information remains accessible and that the signature stays linked to the signed record. A successful vendor demonstration does not establish that the customer's configuration behaves the same way. Failed tests need an assessed resolution, not deletion from the validation report.

What changes when a system is replaced?

A migration needs a defined source of truth and acceptance criteria for the records being moved. Compare representative content, attachments, relationships and signature history, and decide how the original records remain available. Keep the migration evidence and unresolved exceptions. Do not turn historical approvals into fresh signatures by importing the final status alone.

Plan account removal when a person leaves without erasing their historical identity. Reassess changes to permissions, signature workflows, exports and storage against intended use. The organization remains responsible for procedures and trained users; software controls do not authorize someone to sign on another person's behalf.

This guide is an orientation, not regulatory advice. Read 21 CFR Part 11 and the FDA's Part 11 scope-and-application guidance directly, and confirm your implementation with your quality and regulatory team.