Guide
Design controls for medical devices: a practical guide
Design controls are the documented process that turns user needs into a safe, effective device you can manufacture. ISO 13485:2016 clause 7.3 describes them, and the FDA QMSR applies them through that standard. The result is a design history that proves every requirement was met.
The stages
From user needs to production
Planning and inputs
A design plan sets stages, reviews and responsibilities. Design inputs turn user needs into clear, testable requirements, including regulatory and risk requirements.
Outputs and reviews
Outputs are the drawings, specifications and software that define the device. Formal design reviews at planned stages check progress.
Verification and validation
Verification shows outputs meet inputs. Validation shows the finished device meets user needs in its intended use, including usability and clinical evaluation where needed.
Transfer and changes
Design transfer moves the design into production. Design changes are reviewed, verified or validated, and approved before they are implemented.
Doing it well
Habits that pass audits
- 1Start the design plan early and update it as the project changes.
- 2Write inputs that are measurable, so each one has a clear verification method.
- 3Keep a traceability matrix linking needs, inputs, outputs, tests and risk controls.
- 4Tie design reviews to decisions, with an independent reviewer and signed records.
- 5Run design changes through change control so the design history stays complete.
Where Qevatrix fits
Design history with traceability built in
QualityOS keeps design plans, inputs, outputs, reviews, verification, validation and changes linked in one record, with Part 11 signatures and a rule that nobody approves their own work.
How do you show that a requirement has been verified?
Link the requirement to the relevant output, approved verification method and result, then record the review decision. A green status or attached file alone does not demonstrate that the acceptance criteria were met. Failed tests and unresolved gaps must remain visible until assessed through the controlled process.
What does a practical example look like?
Illustrative example: a manual surgical tool has a dimensional requirement. The team links the drawing version, test method and measured result to that requirement and reviews the specified acceptance criteria. If the test fails, the record stays open while the design or evidence is assessed; a completion indicator never approves the design.
What should the reviewer check before proceeding?
- Use measurable acceptance criteria tied to the requirement.
- Record output and test versions, results and reviewer decisions.
- Keep verification, validation and transfer decisions distinct.
Where can you check the original requirements?
Read the primary sources and confirm their applicability to your device and market. This example is illustrative, not a classification decision, clinical conclusion or approval.
How do you turn a vague need into a testable input?
Separate what the user needs from the engineering solution. A statement such as 'easy to handle' may describe a need, but it does not yet define how the design will be assessed. Identify the intended user, use environment and relevant risk before writing an input with a justified acceptance criterion. The criterion must come from the device's evidence and responsible specialists, not a generic starter template.
Illustrative traceability chain: a handling need leads to an identified input, a drawing or specification, a verification method and a reviewed result. Assign stable identifiers to those records so revisions can be followed. A single output can address several inputs, and one input can need several evidence sources. Traceability is therefore more than a one-row-per-test spreadsheet; it should expose uncovered requirements and the reasons an item is not applicable.
What does a failed verification mean for the project?
A failed result should remain linked to its method, tested configuration and acceptance criteria. Determine whether the cause is the design, the method, execution or an unresolved investigation. Record the assessment and controlled next step. Retesting without preserving the original result makes the history harder to reconstruct and does not itself resolve the underlying issue.
Consider a hypothetical manual instrument with four inputs, of which three have reviewed verification evidence and one remains unresolved. The arithmetic describes evidence coverage only: it does not establish that the device is safe, validated or ready for transfer. A critical unresolved requirement can block the next stage even when most tasks are complete. The project owner should track both progress and release blockers rather than average them into a compliance score.
How do changes stay connected to transfer?
When an output changes, identify the affected inputs, risk controls, tests, labeling and production information. Review which existing evidence still applies and which work must be repeated. Keep the rationale with the change and preserve the prior approved version. A revised drawing should not automatically overwrite the configuration referenced by a historical test.
Transfer review connects the approved design to the specifications and procedures used in production. Confirm responsibilities where design and manufacturing are split between organizations. Premarket exemption is not a general exemption from every design, quality or identification obligation; the regulatory lead must assess applicability for the device and market. Templates and project boards organize this work but cannot perform the required human review or approval.
More guides
See all guidesQualityOS
FDA QMSR transition guide
What changes as 21 CFR Part 820 aligns with ISO 13485:2016, and how to prepare your quality system.
QualityOS
ISO 13485 in 2026
What is being worked on in the standard, what stays the same, and what to plan for.
QualityOS
Medical device quality management system
The processes a device QMS needs, how they connect, and a practical order to build them in.
This guide is an orientation, not regulatory advice. Confirm requirements against ISO 13485:2016 clause 7.3, the FDA QMSR and the regulations of each market you sell in.
